CHARM Characteristics |
| Parameter | Sample Material | Sample Volume | Assay Range | Precision within run |
|---|---|---|---|---|
| Lipoprotein(a) | Serum or EDTA plasma | 10 µl | 10 - 100 mg/dl | CV < 2.5% (at 50 mg/dl) |
Sample MaterialFresh sample material: 10 μl serum or EDTA plasma. Test KitERS cuvette filled with glycine buffer solution Measurement Range10 - 100 mg/dl Precision within runN = 20; mean = 50 mg/dl; CV = 2.46% CorrelationN = 40; y (Euromedix Lp(a)) = 0.9918 x |